Advertisement

Repeatability and reproducibility of manual choroidal thickness measurement using Lenstar images in children before and after orthokeratology treatment

  • Biyue Guo
    Correspondence
    Corresponding author at: A136-137, Optometry Research Clinic, School of Optometry, The Hong Kong Polytechnic University, Hung Hum, Kowloon, Hong Kong Special Administrative Region.
    Affiliations
    Centre for Myopia Research, School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong Special Administrative Region
    Search for articles by this author
  • Jason K. Lau
    Affiliations
    Centre for Myopia Research, School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong Special Administrative Region
    Search for articles by this author
  • Sin Wan Cheung
    Affiliations
    Centre for Myopia Research, School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong Special Administrative Region
    Search for articles by this author
  • Pauline Cho
    Affiliations
    Centre for Myopia Research, School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong Special Administrative Region
    Search for articles by this author

      Abstract

      Purpose

      To investigate the repeatability and reproducibility of choroidal thickness measurements using Lenstar images in young myopic children before and after one-month orthokeratology (ortho-k) treatment.

      Method

      Ocular biometry of 39 subjects were performed using the Lenstar 900. The first five measurements with maximum differences of 0.02 mm in axial length in the right eyes were saved and used for measurement of choroidal thickness. Subfoveal choroidal thickness were manually measured by identifying the signals from the retinal pigmented epithelium layer and chorioscleral interface. Repeatability was determined by comparing measurements of the same images made by the same observer on two separate occasions (four weeks apart), while reproducibility was calculated by comparing measurements of the same images made by two independent observers. Data was analysed using intra-class correlation coefficients (ICC) and non-parametric Bland and Altman plots.

      Results

      The choroidal peaks could not be identified in all five measurements in all subjects. On average, only 71% subjects had at least four definable images. Compared with the use of fewer than four images, reliability using an average of four definable images improved statistically, but remained clinically unacceptable (>10 µm), although pre- and post-ortho-k ICC values were good to excellent for repeatability (0.867 and 0.975, respectively) and excellent and good for reproducibility (0.959 and 0.868, respectively). Non-parametric pre- and post-ortho-k limits of agreement (2.5% and 97.5% percentiles) obtained were −45.8 to 79.3 µm and −30.3 to 9.5 µm, respectively for repeatability, and −29.0 to 33.5 µm and −21.8 to 70.0 µm, respectively for reproducibility.

      Conclusion

      Choroidal thickness measurements using the Lenstar did not show good reliability, despite the high ICC values, non-parametric Bland and Altman plots demonstrated a wide variability of measurement errors. Any changes in subfoveal choroidal thickness, measured by Lenstar, of <80 µm may not represent real changes.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Contact Lens and Anterior Eye
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Wildsoet C.
        • Wallman J.
        Choroidal and scleral mechanisms of compensation for spectacle lenses in chicks.
        Vision Res. 1995; 35: 1175-1194
        • Troilo D.
        • Nickla D.L.
        • Wildsoet C.F.
        Choroidal thickness changes during altered eye growth and refractive state in a primate.
        Invest Ophthalmol Vis Sci. 2000; 41: 1249-1258
        • Hung L.-F.
        • Crawford M.L.J.
        • Smith E.L.
        Spectacle lenses alter eye growth and the refractive status of young monkeys.
        Nat Med. 1995; 1: 761-765
        • Smith E.L.
        • Hung L.-F.
        • Huang J.
        • Blasdel T.L.
        • Humbird T.L.
        • Bockhorst K.H.
        Effects of optical defocus on refractive development in monkeys: evidence for local, regionally selective mechanisms.
        Invest Ophthalmol Vis Sci. 2010; 51: 3864
        • Mutti D.O.
        Hereditary and environmental contributions to emmetropization and myopia.
        Optom Vis Sci. 2010; 87: 255-259
        • Grosvenor T.
        Reduction in axial length with age: an emmetropizing mechanism for the adult eye?.
        Optom Vis Sci. 1987; 64: 657-663
        • Smith III, E.L.
        • Hung L.F.
        • Huang J.
        • Arumugam B.
        Effects of local myopic defocus on refractive development in monkeys.
        Optom Vis Sci. 2013; 90: 1176
        • Sander B.P.
        • Collins M.J.
        • Read S.A.
        The interaction between homatropine and optical blur on choroidal thickness.
        Ophthalmic Physiol Opt. 2018; 38: 257-265
        • Chakraborty R.
        • Read S.A.
        • Collins M.J.
        Monocular myopic defocus and daily changes in axial length and choroidal thickness of human eyes.
        Exp Eye Res. 2012; 103: 47-54
        • Chakraborty R.
        • Read S.A.
        • Collins M.J.
        Hyperopic defocus and diurnal changes in human choroid and axial length.
        Optom Vis Sci. 2013; 90: 1187-1198
        • Bulut A.
        • Öner V.
        • Büyüktarakçı Ş.
        • Kaim M.
        Associations between choroidal thickness, axial length and spherical equivalent in a paediatric population.
        Clin Exp Optom. 2016; 99: 356-359
        • Read S.A.
        • Collins M.J.
        • Vincent S.J.
        • Alonso-Caneiro D.
        Choroidal thickness in myopic and nonmyopic children assessed with enhanced depth imaging optical coherence tomography.
        Invest Ophthalmol Vis Sci. 2013; 54: 7578https://doi.org/10.1167/iovs.13-12772
        • Read S.A.
        • Alonso-Caneiro D.
        • Vincent S.J.
        • Paul F.
        Longitudinal changes in macular retinal layer thickness in pediatric populations: myopic vs non-myopic eyes.
        PLoS ONE. 2017; 12: e0180462
        • Read S.A.
        • Alonso-Caneiro D.
        • Vincent S.J.
        • Collins M.J.
        Longitudinal changes in choroidal thickness and eye growth in childhood.
        Invest Ophthalmol Vis Sci. 2015; 56 (3103-12)
        • Holden B.A.
        • Fricke T.R.
        • Wilson D.A.
        • Jong M.
        • Naidoo K.S.
        • Sankaridurg P.
        • et al.
        Global prevalence of myopia and high myopia and temporal trends from 2000 through 2050.
        Ophthalmology. 2016; 123: 1036-1042
        • Rudnicka A.R.
        • Kapetanakis V.V.
        • Wathern A.K.
        • Logan N.S.
        • Gilmartin B.
        • Whincup P.H.
        • et al.
        Global variations and time trends in the prevalence of childhood myopia, a systematic review and quantitative meta-analysis: implications for aetiology and early prevention.
        Br J Ophthalmol. 2016; 100: 882-890
        • Ding B.-Y.
        • Shih Y.-F.
        • Lin L.L.K.
        • Hsiao C.K.
        • Wang I.-J.
        Myopia among schoolchildren in East Asia and Singapore.
        Surv Ophthalmol. 2017; 62: 677-697
        • Cho P.
        • Cheung S.-W.
        Retardation of myopia in orthokeratology (ROMIO) study: a 2-year randomized clinical trial.
        Invest Ophthalmol Vis Sci. 2012; 53: 7077https://doi.org/10.1167/iovs.12-10565
        • Walline J.J.
        • Jones L.A.
        • Sinnott L.
        • Manny R.E.
        • Gaume A.
        • Rah M.J.
        • et al.
        A randomized trial of the effect of soft contact lenses on myopia progression in children.
        Invest Ophthalmol Vis Sci. 2008; 49: 4702https://doi.org/10.1167/iovs.08-2067
        • Chia A.
        • Chua W.-H.
        • Cheung Y.-B.
        • Wong W.-L.
        • Lingham A.
        • Fong A.
        • et al.
        Atropine for the treatment of childhood myopia: safety and efficacy of 0.5%, 0.1%, and 0.01% doses (Atropine for the Treatment of Myopia 2).
        Ophthalmology. 2012; 119: 347-354
        • Chen C.
        • Cheung S.W.
        • Cho P.
        Myopia control using toric orthokeratology (TO-SEE study).
        Invest Ophthalmol Vis Sci. 2013; 54: 6510-6517
        • Santodomingo-Rubido J.
        • Villa-Collar C.
        • Gilmartin B.
        • Gutiérrez-Ortega R.
        Myopia control with orthokeratology contact lenses in Spain: refractive and biometric changes.
        Invest Ophthalmol Vis Sci. 2012; 53: 5060https://doi.org/10.1167/iovs.11-8005
        • Binkhorst R.D.
        The accuracy of ultrasonic measurement of the axial length of the eye.
        Ophthalmic Surg Lasers Imaging Retina. 1981; 12: 363-365
        • Olsen T.
        The accuracy of ultrasonic determination of axial length in pseudophakic eyes.
        Acta Ophthalmol. 1989; 67: 141-144
        • Rose L.T.
        • Moshegov C.N.
        Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation.
        Clin Exp Ophthalmol. 2003; 31: 121-124
      1. Barkana Y, Gerber Y, Elbaz U, Schwartz S, Ken-Dror G, Avni I et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 2005;31:1729–35.

        • Huang D.
        • Swanson E.
        • Lin C.
        • Schuman J.
        • Stinson W.
        • Chang W.
        • et al.
        Optical coherence tomography.
        Science. 1991; 254: 1178-1181
        • Dong J.
        • Zhang Y.
        • Zhang H.
        • Jia Z.
        • Zhang S.
        • Wang X.
        • et al.
        Comparison of axial length, anterior chamber depth and intraocular lens power between IOLMaster and ultrasound in normal, long and short eyes.
        PLoS ONE. 2018; 13: e0194273
        • Copete S.
        • Flores-Moreno I.
        • Montero J.A.
        • Duker J.S.
        • Ruiz-Moreno J.M.
        Direct comparison of spectral-domain and swept-source OCT in the measurement of choroidal thickness in normal eyes.
        Br J Ophthalmol. 2014; 98: 334-338
        • Li Z.
        • Cui D.
        • Hu Y.
        • Ao S.
        • Zeng J.
        • Yang X.
        Choroidal thickness and axial length changes in myopic children treated with orthokeratology.
        Cont Lens Anterior Eye. 2017; 40: 417-423
        • Chen Z.
        • Xue F.
        • Zhou J.
        • Qu X.
        • Zhou X.
        Effects of orthokeratology on choroidal thickness and axial length.
        Optom Vis Sci. 2016; 93: 1064-1071
        • Gardner D.J.
        • Walline J.J.
        • Mutti D.O.
        Choroidal thickness and peripheral myopic defocus during orthokeratology.
        Optom Vis Sci. 2015; 92: 579-588
        • Lau J.K.
        • Wan K.
        • Cheung S.-W.
        • Vincent S.J.
        • Cho P.
        Weekly changes in axial length and choroidal thickness in children during and following orthokeratology treatment with different compression factors.
        Transl Vis Sci Technol. 2019; 8: 9
        • Boonarpha N.
        • Zheng Y.
        • Stangos A.N.
        • Lu H.
        • Raj A.
        • Czanner G.
        • et al.
        Standardization of choroidal thickness measurements using enhanced depth imaging optical coherence tomography.
        Int J Ophthalmol. 2015; 8: 484
        • Sim D.A.
        • Keane P.A.
        • Mehta H.
        • Fung S.
        • Zarranz-Ventura J.
        • Fruttiger M.
        • et al.
        Repeatability and reproducibility of choroidal vessel layer measurements in diabetic retinopathy using enhanced depth optical coherence tomography.
        Invest Ophthalmol Vis Sci. 2013; 54 (2893-901)
        • Hanumunthadu D.
        • Ilginis T.
        • Restori M.
        • Sagoo M.
        • Tufail A.
        • Balaggan K.S.
        • et al.
        Spectral-domain optical coherence tomography retinal and choroidal thickness metric repeatability in age-related macular degeneration.
        Am J Ophthalmol. 2016; 166: 154-161
        • Rahman W.
        • Chen F.K.
        • Yeoh J.
        • Patel P.
        • Tufail A.
        • Da Cruz L.
        Repeatability of manual subfoveal choroidal thickness measurements in healthy subjects using the technique of enhanced depth imaging optical coherence tomography.
        Invest Ophthalmol Vis Sci. 2011; 52: 2267-2271
        • Alonso-Caneiro D.
        • Read S.A.
        • Collins M.J.
        Automatic segmentation of choroidal thickness in optical coherence tomography.
        Biomed Opt Express. 2013; 4 (2795-812)
        • Twa M.D.
        • Schulle K.L.
        • Chiu S.J.
        • Farsiu S.
        • Berntsen D.A.
        Validation of macular choroidal thickness measurements from automated SD-OCT image segmentation.
        Optom Vis Sci. 2016; 93: 1387-1398
        • Lau J.K.
        • Cheung S.W.
        • Collins M.J.
        • Cho P.
        Repeatability of choroidal thickness measurements with Spectralis OCT images.
        BMJ Open. 2019; 4: e000237
        • Buckhurst P.J.
        • Wolffsohn J.S.
        • Shah S.
        • Naroo S.A.
        • Davies L.N.
        • Berrow E.J.
        A new optical low coherence reflectometry device for ocular biometry in cataract patients.
        Br J Ophthalmol. 2009; 93: 949-953
        • Cruysberg L.P.J.
        • Doors M.
        • Verbakel F.
        • Berendschot T.T.J.M.
        • De Brabander J.
        • Nuijts R.M.M.A.
        Evaluation of the Lenstar LS 900 non-contact biometer.
        Br J Ophthalmol. 2010; 94: 106-110
        • Holzer M.P.
        • Mamusa M.
        • Auffarth G.U.
        Accuracy of a new partial coherence interferometry analyser for biometric measurements.
        Br J Ophthalmol. 2009; 93: 807-810
        • Read S.A.
        • Collins M.J.
        • Sander B.P.
        Human optical axial length and defocus.
        Invest Ophthalmol Vis Sci. 2010; 51: 6262-6269
        • Read S.A.
        • Collins M.J.
        • Alonso-Caneiro D.
        Validation of optical low coherence reflectometry retinal and choroidal biometry.
        Optom Vis Sci. 2011; 88: 855-863
        • Woodman E.C.
        • Read S.A.
        • Collins M.J.
        Axial length and choroidal thickness changes accompanying prolonged accommodation in myopes and emmetropes.
        Vision Res. 2012; 72: 34-41
        • Wan K.
        • Lau J.K.
        • Cheung S.W.
        • Cho P.
        Refractive and corneal responses of young myopic children to short-term orthokeratology treatment with different compression factors.
        Cont Lens Anterior Eye. 2020; 43: 65-72
      2. Bland JM, Altman DG. Statistics notes: measurement error. Br Med J 1996;313:744.

        • Frey M.E.
        • Petersen H.C.
        • Gerke O.
        Nonparametric limits of agreement for small to moderate sample sizes: a simulation study.
        Stats. 2020; 3: 343-355
        • Read S.A.
        • Collins M.J.
        • Vincent S.J.
        • Alonso-Caneiro D.
        Choroidal thickness in childhood.
        Invest Ophthalmol Vis Sci. 2013; 54: 3586-3593
        • Li Z.
        • Hu Y.
        • Cui D.
        • Long W.
        • He M.
        • Yang X.
        Change in subfoveal choroidal thickness secondary to orthokeratology and its cessation: a predictor for the change in axial length.