If you don't remember your password, you can reset it by entering your email address and clicking the Reset Password button. You will then receive an email that contains a secure link for resetting your password
If the address matches a valid account an email will be sent to __email__ with instructions for resetting your password
Centre for Ocular Research & Education (CORE), School of Optometry and Vision Science, University of Waterloo, Waterloo, Ontario, CanadaCentre for Eye and Vision Research (CEVR), Hong Kong
Contact lens materials have undergone significant changes over the past 20 years, particularly with respect to the introduction of silicone hydrogel materials. Whilst this development addressed hypoxic issues, other important areas relating to contact lens success, notably comfort, require further research.
Contact lens wettability remains a crucially important part of biocompatibility. Contact lenses can be made more wettable by incorporation of surfactants into blister packs, internal wetting agents, surface treatments or care solutions. However, there remains no clear association between contact lens wettability and comfort, making it challenging to determine the potential for these approaches to be of significant clinical benefit. Most contact lenses are used on a daily wear, reusable basis, which requires them to be disinfected when not worn. The ideal disinfecting solution would also improve comfort during wear. However, balancing these requirements with other factors, including biocompatibility, remains a challenge. Soft lens materials invariably take up and subsequently release certain components of disinfecting solutions onto the ocular surface. This may affect tear film stability and the normal ocular microbiome, and further research is needed in this area to determine whether this has any affect on comfort. Finally, contact lens materials sorb components of the tear film, and these interactions are complex and may change the biochemistry of the tear film, which in turn may affect their comfort.
In conclusion, the interaction between lens materials, tear film and disinfection solution plays an important role in the biocompatibility of lenses. However, the exact role and whether this can be altered to improve biocompatibility and comfort during wear remains debatable. This report summarises the best available evidence to examine this complex relationship and the opportunities for practitioners to enhance in-eye comfort of contemporary lenses, along with providing suggestions for areas of study that may provide further information on this topic.
poly-2-ethyl-2-oxazoline, hydroxypropyl methylcellulose (also known as hypromellose)
I-309
also known as chemokine (C-C motif) ligand 1
ICAM
intercellular adhesion molecule
IFN
interferon
IL
interleukin
ISO
International Organization for Standardization
LTB4
leukotriene B4
LTC4
leukotriene C4
LTD4
leukotriene D4
LTE4
leukotriene E4
MAPD
myristamidopropyl dimethylamine
MCP1
monocyte chemoattractant protein 1
MMP-9
matrix metalloproteinase 9
NIBUT
non-invasive tear break-up time
NVP
n vinyl pyrrolidone
PBS
phosphate buffered saline
PDMS
poly(dimethylsiloxane) dialkanol
PEG
polyethylene glycol (also known as polyethylene oxide or polyoxyethylene depending on its molecular weight)
PEM(s)
polyelectrolyte multilayer(s)
PGE
prostaglandin E
PGF
prostaglandin F
PLTF
pre-lens tear film
pMPC
poly(2-methacryloyloxyethyl phosphorylcholine)
PQ-1
polyquaternium-1
PRG4
proteoglycan 4
PVA
polyvinyl alcohol
PVP
polymer polyvinyl pyrrolidone (also known as povidone)
θ
contact angle
θa
advancing contact angle
θr
receding contact angle
SPEED
Standardised Patient Evaluation of Eye Dryness
TFOS
Tear Film and Ocular Surface Society
TIMP(s)
tissue inhibitor of metalloproteinase(s)
TRIS
tris-(trimethyl-silyl-propyl-methacrylate
v/v
volume by volume
1. Introduction
Contact lenses are medical devices that are primarily used for the correction of ametropia. The vast majority of today’s market, which accounts for approximately US$9 billion worldwide, is made up of soft contact lenses (∼90 %; Table 1), while rigid lenses make up the remainder [
]. This report outlines various changes in polymer and surface chemistry of contact lenses and how these result in improved oxygen permeability, wettability and interactions with the tear film. The report also discusses the use of disinfecting solutions and how these interact with the lens materials.
Table 1Classification of soft contact lens materials according to the International Organization for Standardization, ISO18369-1: 2017 [
Materials having oxygen permeability (Dk) greater than 40 Dk units and that have a Dk greater than that expected on the basis of the material’s water content alone.
V A
Ionic subgroup
A subgroup of 5 which contains monomers or oligomers which are ionic at pH 6 to pH 8
V B
High water subgroup
A subgroup of 5 which contains 50 % water or more and no ionic monomer or oligomer at pH 6 to pH 8
V C
Low water subgroup
A subgroup of 5 which contains less than 50 % water and no ionic monomer or oligomer at pH 6 to pH 8
The cornea is an avascular tissue and consequently obtains its oxygen supply from the oxygen dissolved in the tear film from the air. Polymethyl methacrylate (PMMA) was the first polymer used for manufacturing rigid contact lenses, but had insufficient oxygen permeability, which was the main cause of adverse ocular events [
]. Its modified version (2-hydroxyethyl methacrylate; HEMA) was used to create soft, flexible hydrogel lenses, which were composed of hydrophilic monomers that allowed for interaction with water [
O. Wichterle. Method of manufacturing soft and flexible contact lenses. Google Patents. US3496254A. USA: Czech Academy of Sciences CAS; 1965:US 3498254.
]. HEMA absorbs a large percentage of water within its polymer network due to the presence of highly electronegative atoms such as oxygen. However, there was a need for different types of polymers to further increase the oxygen permeability of lenses to overcome ocular complications such as corneal swelling, conjunctival and limbal hyperaemia and the effects of depleted oxygen on corneal endothelial cells [
]. This section will not cover the response of the ocular surface cells and tissues to oxygen and the benefits of higher oxygen permeable lenses, as this is covered in the CLEAR Complications Report and CLEAR Sclerals Report [
]. Coulometric methods use a nitrogen carrier gas to transport oxygen that has permeated through a lens to an oxygen detector. With the polarographic method, the oxygen is detected by an electrode which is in direct contact with the posterior lens surface.
Oxygen permeability of contact lens materials is mostly attributed to the incorporation of silicon (and sometimes fluorine) into contact lens polymers. When silicon is combined with oxygen in siloxane, the silicon-oxygen bonds allows the transmission of oxygen through the polymers [
]. Siloxane, in the form of silicones (a polymer of siloxane), are now incorporated into many contact lens polymers, forming the basis for silicone hydrogel (SiHy) contact lenses and is incorporated into rigid corneal lenses. Fluorocarbons can dissolve oxygen [
] and these molecules have also been incorporated into contact lens polymers. Whilst this ability to improve characteristics of oxygen permeability has undoubtedly been important for ocular surface biocompatibility and physiology [
], and this has led to issues related to deposition of tear film components onto lenses and interactions with microbes. Although the increase in hydrophobicity can be partly mitigated by surface modifications or incorporation of wetting agents, the characteristics of silicone (or fluorine) containing polymers for contact lenses are still different to their hydrogel counterparts.
2.1 Deposition of tear film proteins and lipids
From a biochemical point of view, the main issue with the incorporation of silicone (or fluorine) into lenses is the change in deposition of tear film components onto lenses [
]. For a SiHy contact lens based on tris-(trimethyl-silyl-propyl-methacrylate) [TRIS]- N, N-dimethyl acrylamide [DMA]-n vinyl pyrrolidone [NVP]- HEMA, increasing the TRIS (silicone) content of the polymers reduced the adsorption of either albumin or lysozyme [
]. As reviewed recently, lysozyme tends to adsorb to hydrogels (especially group IV) in higher quantities than to SiHys either in vitro or during wear (ex vivo analyses) [
]. Silicone hydrogel lenses (lotrafilcon A and balafilcon A) adsorb less protein from an artificial tear solution than hydrogel lenses (polymacon, nelfilcon A or etafilcon A) [
]. The amount of total lipid, cholesteryl esters, cholesterol and triglycerides/phospholipids is greater on balafilcon A SiHys after wear than on etafilcon A hydrogel lenses [
Protein deposition on either hydrogel (polymacon, nelfilcon A or etafilcon A) or SiHy (lotrafilcon A and balafilcon A) lenses produces a small but measurable decrease in the oxygen transmissibility of the lenses [
], studies have been conducted to see if additions to the polymer matrix of SiHy lenses can reduce protein adsorption. Adding 2-methacryloyloxyethyl phosphorylcholine when synthesising SiHys increases their hydrophilicity and reduces the adsorption of albumin [
]. The addition of poly(ethylene glycol) methacrylate during synthesis of SiHys also increases their wettability and decreases adsorption of albumin and lysozyme [
]. Bacterial adhesion to so-called first generation SiHy lenses (lotrafilcon A, lotrafilcon B and balafilcon A) is often greater than to etafilcon A hydrogel lenses [
Effects of quorum sensing molecules of Pseudomonas aeruginosa on organism growth, elastase B production, and primary adhesion to hydrogel contact lenses.
In vitro deposition of lysozyme on etafilcon a and balafilcon a hydrogel contact lenses: effects on adhesion and survival of Pseudomonas aeruginosa and Staphylococcus aureus.
], but there is commonly no difference in adhesion of bacteria between etafilcon A and nelfilcon A hydrogel lenses and later generations of SiHy lenses such as galyfilcon A or senofilcon A [
Interestingly, these in vitro adhesion results do not necessarily translate to the numbers of microbes isolated from contact lenses during wear. The frequency of isolation of the normal microbiota of the ocular surface from contact lenses is very similar between hydrogel and SiHy lenses [
]. Notwithstanding the latter, the increased adhesion of bacteria to the first generation SiHy contact lenses in laboratory studies may be a contributing factor to the increased risk of developing a corneal infiltrative event during wear of SiHys [
Age and other risk factors for corneal infiltrative and inflammatory events in young soft contact lens wearers from the contact lens assessment in youth (clay) study.
There have been several investigations into ways of minimising microbial adhesion to contact lenses and most of these will be discussed in later sections, as they are not related to oxygen permeability. However, two studies have examined whether changing the oxygen permeability of contact lens polymers might, albeit perhaps serendipitously, alter bacterial adhesion. A series of SiHy lens polymers was produced where the silicone monomer to HEMA ratio or the DMA:HEMA ratio was altered [
]. Changes to the silicone monomer to HEMA ratio did not alter the adhesion of Staphylococcus aureus to the materials, whereas increasing the ratio of DMA did decrease bacterial adhesion [
], and this was related to increases in the water content of the lenses. When boric acid is incorporated into HEMA lenses there is an increase in oxygen permeability, driven most likely by an increase in water content [
]. There was also an increase in surface hydrophilicity. As the content of boric acid increased there was an increase in the ability of the hydrogels to kill strains of Escherichia coli, S. aureus, Pseudomonas aeruginosa and Enterococcus faecium [
]. However, as the bacterial killing was measured by a diffusion process, it would appear that there was leakage of an agent (probably boric acid or borate) from the polymers and so the antibacterial nature of the polymers per se was not proven.
2.3 Summary
Increases in the oxygen permeability of contact lenses has brought with it changes to the types of tear film components adsorbed to lenses (proteins vs. lipids) and also affected microbial adhesion (at least in laboratory studies). These changes may be driven by the inherent hydrophobicity of the polymers used to increase oxygen permeability. As the biocompatibility of contact lenses is associated with their surface wettability, efforts have been made to improve the wettability of contact lenses.
3. Achieving wettability
3.1 Definition of wettability
Wetting is the ability of a liquid deposited on a solid surface (or the surface of another immiscible liquid) to spread out and maintain contact with that surface [
]. When a contact lens is placed on the eye, the tear film and the air compete for the contact lens surface and the extent of the surface that is covered by the tear film reflects the wettability of the specific contact lens material. Indeed, wetting involves the three phases of matter: vapor (typically air), liquid, and solid. To understand the interaction between these three phases, the surface free energy of the solid and the surface tension of the liquid have to be taken into consideration. Wettability is promoted by a relatively high surface free energy of the solid and a relatively low surface tension of the liquid.
The surface free energy of a solid is the excess energy at the surface of a material compared to the bulk [
]. It corresponds to the energy of the intermolecular bonds that are broken when a surface is created, i.e. it corresponds to the work required to create the surface from the bulk. Wettability is promoted by a relatively high surface free energy of the solid, such as in the case of glass, where covalent bonds are broken to create the surface. In the case of molecular materials such as polymers made of hydrocarbons, the molecules are often held together by weaker van der Waals and hydrogen bonds. This means that lower energy is needed to create the surface and wetting is not favoured. In this case, wetting depends mostly on the properties of the liquid.
The surface tension of a liquid is conceptually the analogue of surface free energy for solids and is the net energy of the attraction of molecules to each other, i.e. it represents the tendency of the molecules of a liquid to adhere to each other [
]. The attraction of liquid molecules to each other minimises the surface exposed to air, and so the liquid tends to have a spherical shape. A relatively low surface tension of liquid aids its ability to wet a solid surface. Water has a relatively high surface tension (72.8 10−3 J/m2 at 20 °C in SI units corresponding to 72.8 dynes/cm in the centimetre–gram–second system) [
] and this, combined with the low surface free energy of polymers, reduces wettability of the polymers in water.
The wettability of a surface can be modified by the deposition or incorporation of various molecular species on the solid surface or in the liquid. Wetting agents are called ‘surfactants’ (surface active agents). Surfactants increase the spreading and penetrating properties of a liquid by lowering its surface tension. The wettability of contact lens materials depends on the chemistry of the contact lens constituents, manufacturing processes, surface coatings, and addition of internal or releasable wetting agents. In the case of reusable contact lenses, initial wettability out of the blister pack may be modified by the solutions used for daily lens care and the deposition of tear film components onto the lenses.
3.2 Wettability measurement methods
Experimental methods to assess the wettability of contact lens materials can be laboratory-based, using unworn (in vitro) or worn (ex vivo) lenses, or those that assess wettability on the eye (in vivo clinical assessments).
A common method for assessing wettability of a liquid deposited on a contact lens, either in vitro or ex vivo, is the measurement of the contact angle [
]. The contact angle is the angle between the substrate surface and the tangent drawn at the triple point between the three phases where the liquid–vapor interface meets the solid-liquid interface (θ in Fig. 1) [
In the ideal case of a pure liquid spreading on a smooth inert solid, Young's equation gives the equilibrium contact angle in terms of interfacial tensions existing at the three-phase interface [
where γ is the interface tension between two phases at the interface which can be the solid (S), the liquid (L), or the vapor (V) phase, as indicated by the subscripts in Eq. (1). When the vapor phase is air, the interface tension is the surface free energy of the solid and the interface tension is the surface tension of the liquid. The difference ( is the adhesion tension, which describes the liquid attraction towards the solid. As can be deduced from Eq. (1), the degree of wetting results from the balance between the intermolecular adhesive interactions represented by the numerator (liquid to surface interactions, causing a liquid drop to spread out on the surface) and the cohesive interactions represented by the denominator (liquid to liquid, causing the drop to ‘ball up’ to avoid the contact with the surface) [
]. The θ angle, which is between 0° and 180°, is less than 90° (wettable surface) if the adhesive interaction at the numerator is positive because, in this case, the cosine of the angle is positive. On the contrary, when the cosine of the angle is negative (because the adhesive interaction at the numerator is negative), the cosine of the angle is negative and, thus, the surface is non-wettable (θ > 90°). Wettability is maximum when the adhesive interaction at the numerator of Eq. (1) tends to the denominator value. In this case, the cosine of the angle tends to 1 and, therefore, the θ angle tends to 0°.
An ideal surface is flat, rigid, smooth, and chemically homogeneous. In this ideal case, a relatively simple measurement of the static contact angle can be made. Table 2 summarises the techniques adopted in the literature to measure contact angles and wettability of contact lenses.
Table 2In vitro methods to assess the surface wettability of contact lenses.
The main static methods to measure wettability are:
•
Sessile drop goniometry (Fig. 2). A droplet of liquid is deposited by a syringe on a solid surface. A digital camera and a goniometer are used to capture the profile and measure the contact angle. In the case of polymers, especially for hydrated samples, the measurements must be performed in a temperature- and humidity-controlled environment and in a relatively short time, since the drop permeation into the matrix of the material can lead to variability of the measured angle or the sample itself can dehydrate.
Fig. 2Diagram showing the contact angle (θ) measured by the static sessile drop method and by the static captive bubble method. S, L, and V indicate the solid, liquid, and vapor phases, respectively.
Captive bubble method (Fig. 2). A sample of the solid of interest is immersed in a liquid. An air bubble or a bubble of another immiscible fluid is injected under the solid surface, i.e. attached from below to the solid surface. The shape of the bubble is then evaluated. This method is often used with hydrogels such as contact lenses, as the sample does not dehydrate.
Contact angles measured by sessile drop are typically less repeatable than when using the captive bubble method, especially for SiHy contact lenses, probably due to surface dehydration and the blotting required to remove excess liquid [
In practice, real surfaces often are not flat, rigid, smooth, or chemically homogeneous, and the measured contact angle may vary from point to point on the surface. Wettability is very sensitive to surface contamination, non-homogeneity and the roughness of the surface. In these cases, the contact angle measurement through a static method is affected by the local slope of the surface and the local chemical composition [
]. In the case of soft contact lenses, sample swelling and deformation of the surface can also influence static measurements of the contact angle. In these cases, measurements of the hysteresis, defined as the difference between the so-called ‘advancing’ and ‘receding’ contact angles, can be useful. Static measurements on non-ideal samples yield values between the advancing and receding contact angles. The hysteresis can also be estimated by comparing results from two methodologies, with the sessile drop method producing the advancing-type angle and the captive bubble method producing the receding-type angle [
Dynamic sessile drop on inclined substrates. A drop is deposited on a substrate and the substrate is inclined. At a given angle of tilt, the advancing and receding contact angles are reached on the two sides of the drop respectively, just before the drop starts moving on the surface.
2
Dynamic sessile drop goniometry (also known as the extension and contraction technique (Fig. 3)). The advancing contact angle is measured by adding liquid to gradually increase the sessile drop in volume. The increase in volume causes an increase of the contact angle without increasing the solid–liquid interfacial area until a maximum angle is reached, and the contact line begins to move across the surface (advancing phase). The receding angle is measured by removing liquid from the drop. The decrease in volume causes a decrease of the observed contact angle without increasing the solid–liquid interfacial area until a stable value of the angle is reached, and the contact line begins to move (receding phase).
Fig. 3Diagram showing the dynamic sessile drop goniometry method (extension and contraction technique) to measure the advancing (θa) and receding (θr) contact angles. S, L, and V indicate the solid, liquid, and vapor phases, respectively.
Dynamic captive bubble method (an extension and contraction technique). The advancing and receding contact angles are measured by gradually expanding and contracting an air bubble in contact with the solid surface (see the static captive bubble method), similarly to the extension and contraction technique based on the sessile drop discussed above and illustrated in Fig. 3.
4
Wilhelmy plate. The solid sample is oriented perpendicularly to the air/liquid interface and it is attached to a balance positioned above the liquid. The wetting force on the solid is measured as the solid is immersed into the liquid (advancing contact angle) or withdrawn back to the initial position (receding contact angle). The measured force is used to calculate the two angles.
In the case of dynamic contact angle measurements on contact lenses, the advancing contact angle is considered to be the parameter most closely describing the initial spreading of the pre-lens tear film (PLTF) on the contact lens surface, whereas the receding contact angle is considered as an indicator of the PLTF stability when the eyelids open and the PLTF starts to retract [
]. Reducing hysteresis, the difference between the advancing and receding contact angles, has been one of the goals in contact lens development in order to produce a more wettable surface [
Other techniques for assessing contact lens wettability in vitro include the observation and characterization of images taken of the tear film over the contact lens. Taking inspiration from some in vivo investigation techniques, a thin film interferometer has been used to capture images of the pre-lens liquid film of contact lenses to describe their drying properties [
]. Images are obtained from the lens surface when it is wet until it becomes dry and the following parameters can be deduced: time of the first break-up (onset latency), duration of lens surface drying (drying duration), maximum speed of increase in the drying area (maximum speed), and the time to reach maximum drying speed (peak latency). A slightly different technique, non-invasive keratograph dry-up time, has shown that dewetting of SiHy contact lenses was non-uniform across the contact lens surface [
]. In general, a more complete characterization of the surface energy may be necessary to better model the behaviour of contact lenses in vivo, rather than just relying on the assessment of the contact angles under specific conditions [
Regarding wetting measurements on contact lenses, one aspect to consider is that soft contact lenses are distributed in blister packs filled with a liquid which may contain buffered sodium chloride solution, preserving agents, disinfecting agents and additives introduced to improve wettability. These components can penetrate inside the porous lens matrix and be released when the lens is exposed to another liquid. The latter can also occur during wettability measurements if the lens is not properly rinsed. Often wettability measurements are carried out in vitro using water or saline. However, a contact lens is wet by tears during wear, and water and saline do not mimic the tear film. In normal participants, tears have a surface tension which is approximately two-thirds that of water or saline (42−46 10−3 J/m2 in SI units corresponding to 42−46 dynes/cm in the centimetre–gram–second system) [
]. The use of saline or water that do not contain these components are unlikely to adequately mimic human tears. In vitro NIBUT measurements of SiHy contact lenses taken immediately out of their blister packs or exposed to an artificial tear solution containing various lipids (i.e. oleic acid methyl ester, cholesterol, triolein, phosphatidylcholine, cholesteryl oleate, and oleic acid), salts, urea, glucose, proteins (lysozyme and hen egg albumin), and mucin (with concentrations based on those in normal human tears) demonstrated that these artificial tears tended to reduce differences in surface wettability [
]. Decorating the surface of a SiHy lens with simple model lipids such as dipalmitoylphosphatidylcholine and cholesterol increased its hydrophilicity and inhibited dewetting, whereas addition of whole meibum reduced hydrophilicity and wetting [
Although this section deals with wettability, it should be noted that materials, manufacturing processes, surface coatings to increase wettability, and wetting agents (internal or released) also contribute to the tribological characteristics of contact lenses, which are expected to play an important role during wear (including affecting comfort of lenses – see section 2.4) [
]. Tribology is the study of the interaction between surfaces in relative motion. In the case of contact lenses, the surfaces in relative motion are those of the corneal surface, the tear film, the contact lens and the eyelid. Friction contributes to the tribological properties of this system and is the force resisting the relative motion of these components sliding against each other. Lubricants help in reducing friction and wetting agents are also lubricants.
3.2.2 Measuring wettability in vivo
When a contact lens is placed onto the ocular surface, factors in the ocular environment such as the temperature, osmolarity and composition of the tears can impact the chemistry of the material, changing its surface properties and in turn wettability [
]. Some attempts to adapt in vitro techniques (Table 2) to assess in vivo wettability have been sporadically proposed in the literature. A measure of contact angle in vivo has been proposed both for rigid corneal lenses [
]. However, these methods need validation before they can be translated into clinical practice.
Indirectly, contact lens wettability can be evaluated by assessing features of the PLTF that provide useful information about the level of tears spreading on the lens. Tear coverage can be assessed by examining any PLTF deficiency on the contact lens surface through slit lamp observation by examining the quality of a specular reflection at high magnification, using interferometry observation techniques and with the use of grading scales [
]. Another possibility is to evaluate the optical quality of the contact lens on the eye in terms of higher-order aberrations, since the surface wetting of a contact lens influences optical quality [
]. Finally, examining the loss of superficial optical quality by measuring the time elapsed between cessation of blinking and blur-out of a threshold letter on the acuity chart has been advanced as another measure of wettability [
However, NIBUT (sometimes called non-invasive surface drying time when used with contact lenses) is the most utilised in vivo method for assessing contact lens wettability. This is performed using an image (keratoscopy, videokeratoscopy, grid, etc.) projected onto the contact lens and evaluating the quality of the reflection from the PLTF [
]. NIBUT of the PLTF can also provide contact lens practitioners with an indirect assessment of the lubricity and thus potentially on-eye friction of the contact lens, which is impossible to measure directly in the eye [
]. NIBUT has several advantages such as accessibility for clinicians, coverage of a large portion of the contact lens surface, and minimum influence of eye movements [
]. NIBUT has been extensively reviewed in the TFOS (Tear Film and Ocular Surface) International Workshop on Contact Lens Discomfort: Report of the Contact Lens Interactions With the Tear Film Subcommittee and TFOS DEWS II Tear Film Reports [
3.3 Material chemistry, manufacturing processes, surface coatings and internal wetting agents
A moist contact lens surface is considered relevant for ocular physiology and comfort, prevention of deposits and optical clarity. A hydrophobic contact lens surface repels tears, whereas hydrophilic surfaces allow moisture to be retained on the lens surface. Silicone hydrogels are among the most popular materials used today [
] unless they are surface treated or copolymerised with other components. Even then, the migration of hydrophobic siloxane components to the surface can cause reduced wettability [
]. All types of contact lenses can be manufactured with wetting agents which may be embedded into the contact lens (internal wetting agents), embedded on its surface (surface wetting agents), or may be progressively released from the contact lens during wear (released wetting agents) to improve their “wearability”. Some common examples of molecules used as contact lens wetting agents are methacrylic acid; glycidyl methacrylate (the ester of methacrylic acid and glycidol), NVP and the corresponding polymer polyvinyl pyrrolidone (PVP; also known as povidone), polyvinyl alcohol (PVA), polyethylene glycol (PEG; also known as polyethylene oxide or polyoxyethylene depending on its molecular weight), poly-2-ethyl-2-oxazoline, hydroxypropyl methylcellulose (HPMC; also known as hypromellose), hyaluronic acid (HA), phosphorylcholine and the phospholipid phosphatidylcholine and various poloxamers [
HEMA-based hydrogel lens materials often incorporate more hydrophilic monomers such as methacrylic acid, GMA (glyceryl methacrylate), NVP, PVA and phosphorylcholine to increase the material water content, thereby increasing material Dk [
]. For example, when HEMA-based hydrogels are modified with NVP, MA or GMA comonomers, the contact angle decreases as the concentration of the crosslinker in the hydrogels increases [
]. Incorporation of surfactants in the manufacture of contact lens materials reduces the contact angle and can reduce the coefficient of friction of hydrogel lenses [
]. A correlation between the contact angle and coefficient of friction might be attributed to the stretching of the surfactant tails near the surface into the aqueous phase, and not only to the general increase in water content.
Photo-crosslinkable methacrylated HA has been incorporated into hydrogel contact lenses as an internal wetting agent [
] and this improved hydrophilicity. HA of lower molecular weight and degree of methacrylation has been associated with an increased mobility and improved hydrophilicity than the less mobile HA [
]. Hydrophilic functionalised silicone-based macromers have also been used with hydrophilic monomers in materials such as comfilcon A, enfilcon A and somofilcon A [
]. Increasing PEG methacrylate in a silicone-based hydrogel formed from poly(dimethylsiloxane) dialkanol (PDMS), isophorone diisocyanate and HEMA led to a lower water contact angle and higher water content [
]. These materials require surface modification to increase their wettability using chemical, mechanical, laser, plasma, electron beam irradiation or micro-structuring of the lens surface (a lotus leaf eff ; ;ect) [
]. The incorporation of nitrogen or oxygen-containing groups on the surface and the transformation of silicone into hydrophilic silicate after plasma treatment are the main reasons for the improvement in surface hydrophilicity [
]. Oxygen plasma has also been employed to improve surface hydrophilicity through the incorporation of oxygen and the transformation of Si CH3 into hydrophilic Si O [
The surface of HEMA-based soft lenses can be modified with the use of wetting agents to improve the hydrophilicity of the lens surface. The first commercial lens material to utilise this concept related to the incorporation of non-releasable PVP into the etafilcon A hydrogel material, which resulted in improved wettability and comfort compared with the base etafilcon A material [
]. Functionalizing the surface of HEMA-based contact lenses, using various methods including "click" chemistry, with HA increases surface wettability and water retention [
]. A common reaction employed in click chemistry is copper-catalyzed azide-alkyne cycloaddition which tolerates a broad range of temperatures, works in aqueous conditions and over a very broad (4–12) pH range [
]. This reaction can proceed by functionalising a surface with an azide coating (or similar) and adding a compound that has been functionalised to contain an alkyne; copper catalyses the formation of a covalent linkage via a 1,3 cycloaddition. HEMA materials treated with sulfonated poly(ethylene glycol) via crosslinking reduced dynamic contact angles and protein adsorption with the increasing concentration of sulfonated poly(ethylene glycol) [
The first commercially available SiHys, balafilcon A and lotrafilcon A, were plasma treated to improve wettability. The surface of balafilcon A was oxidised by plasma [
]. However, clinical studies have shown reduced surface wettability of lotrafilcon A contact lenses compared to the later commercialised (but no longer available) galyfilcon A SiHy material, which was not surface treated, but contained PVP as an internal wetting agent [
A more recent development is the commercialisation of a SiHy contact lens with a water-gradient structure (delefilcon A). This lens is made of a SiHy core and a surrounding hydrogel surface layer a few micrometres thick, with a water content of approximately 80 % [
]. The advancing and receding water contact angles of narafilcon A and senofilcon A (having no surface coating), stenfilcon A (with just 4.4 % bulk silicone content), and delefilcon A (>80 % water at its surface) have been compared using the advancing contact angle [
]. The clinical performance of water gradient delefilcon A lenses has also been shown to be superior in a cross-over, randomised, masked study in comparison to two other daily disposable SiHy lenses (somofilcon A and narafilcon A) [
]. Also, delefilcon A lenses had a longer NIBUT, comparatively greater inferior tear meniscus height and less corneal staining after 16 h of lens wear than narafilcon A and somofilcon A lenses, although no significant differences were found between lens types in subjective comfort, visual acuity, or quality of vision [
]. Novel amphiphilic (both hydrophilic and lipophilic) networks based on polyallyl-methacrylate and PEG with l-cysteine conjugated to the surface have been produced, which could be potentially used to improve lens wettability [
]. The co-networks had relatively low water contact angles (less than 80°, down to 25° for some of these networks) and the l-cysteine improved the surface wettability of these systems [
]. Surface polyelectrolyte multilayers (PEMs) have also been proposed to enhance lens wettability. A SiHy composed of PDMS and PEG methacrylate with a surface PEM of chitosan and HA (as positive- and negative-charged agents, respectively) showed decreased contact angles as the number of PEM grafting layers increased [
Hyaluronan incorporation into model contact lens hydrogels as a built-in lubricant: effect of hydrogel composition and proteoglycan 4 as a lubricant in solution.
]. A different surface modification strategy has also been proposed, whereby a HA-binding peptide is bound to the surface of contact lenses. Bound HA, via the HA-binding peptide, significantly reduced water loss from the modified contact lens [
A dimethyl acrylamide hydrogel layer (2.1 ± 1.4 μm thick) has been coated onto narafilcon A SiHy contact lenses. This resulted in a significant reduction in contact angle from ∼95 degrees for untreated lenses to ∼15 degrees [
]. Nelfilcon A contact lenses are made of a PVA–based hydrogel. Non-functionalised high molecular weight PVA can also be added to contact lenses to promote the release of PVA [