Volume 30, Issue 1, Pages 17-21 (March 2007)


View previous. 6 of 14 View next.

Factors influencing the measurement of oxygen shortfall of the human cornea: Sequencing of test conditions

Jane M. Chi, Barbara A. FinkCorresponding Author Informationemail address, Richard M. Hill, G. Lynn Mitchell

published online 26 November 2006.

Abstract 

Purpose

The effects of sequencing of test conditions, in this case contact lens thicknesses, on the measurement of the oxygen shortfall of human corneas were studied.

Methods

Corneal oxygen uptake rates were measured with a Clark-type polarographic electrode on the central, unanesthetized right corneas of 14 human subjects. Measurements were made under the following conditions: (1) the normal open eye; (2) after 5min of static (without blinking) wear of each of seven rigid gas permeable lenses of seven center thicknesses (0.18, 0.12, 0.16, 0.20, 0.24, 0.28, and 0.32mm); (3) after 5min of static wear of a polymethylmethacrylate (PMMA) contact lens. Lens thicknesses were randomly assigned numbers, which were sequenced in seven cycles. Two subjects were assigned to each sequencing cycle, and each subject participated in two identical sessions.

Results

The interaction of order x thickness was determined to be insignificant (F=0.99; p=0.5101). The effect of lens order was also insignificant (F=0.76; p=0.6239), indicating that the order of lens placement did not affect the measured corneal oxygen shortfall. Not surprisingly, the analysis indicated a significant effect of lens thickness on corneal oxygen shortfall (F=3.94; p=0.0032).

Conclusions

The sequencing of lenses of various thicknesses on the cornea does not affect the measurement of corneal oxygen shortfall.

Article Outline

Abstract

1. Introduction

2. Methods

2.1. Subjects

2.2. Contact lenses

2.3. Procedure

3. Statistical analyses

4. Results

5. Discussion

References

Copyright

1. Introduction 

return to Article Outline

In order to determine the effects of contact lens thickness and transmissibility on the oxygen supply to the cornea, corneal oxygen uptake measurements can be made immediately following the wear of contact lenses for a short period of time [1], [2]. Many factors influence the values that are obtained in these measurements. Procedural factors include: the reservoir of oxygen contained in the polarographic electrode [3], [4], [5], [6], [7], [8], the membrane material and thickness [3], [9], [10], [11], [12], [13], the oxygen tension of the surrounding corneal environment [14], the duration of exposure to reduced oxygen tension [4], [15], [16], [17], [18], the time between exposure and measurement [3], [4], [10], [19], and use of topical anesthetic [20], [21], [22]. Subject factors that influence the measured values of corneal oxygen uptake include, but are not limited to, the following: the influence of contact lens wear on the fellow eye [23], the temperature surrounding the cornea [24], [25], [26], lid position [27], [28], [29], corneal location [30], [31], [32], [33], [34], [35], corneal thickness [3], [34], [35], [36], [37], [38], and corneal health [39], [40], [41].

Contact lens wear results in increased corneal oxygen uptake, particularly if the contact lens has low oxygen transmissibility or if tear exchange is not taking place [42]. The wear of polymethylmethacrylate (PMMA) contact lenses results in an increase in oxygen uptake to about 5–6× that of the normal open eye within about 120s [4], [16]. Corneal oxygen uptake increases less dramatically following the wear of oxygen permeable materials, and at higher contact lens transmissibilities the reduction in corneal oxygen uptake with increasing transmissibility diminishes [1], [2]. Several studies have investigated the effects of contact lens materials, designs, hydration, and movement on corneal oxygen uptake [1], [2], [17], [20], [31], [35], [43], [44], [45], [46], [47], [48], [49], [50], [51], [52], [53], [54], [55], [56], [57], [58], [59].

Four methods are typically used to record and present the cornea's rapid response to the oxygen levels in its environment: (1) absolute units of mmHg/s, (2) as a ratio of corneal response to a test environment and response to air, (3) as a percentage of the cornea's response to hypoxia, such as can occur by exposure to a goggle with 100% nitrogen or by the wear of polymethylmethacrylate contact lenses for 5min, and (4) as a proportion of the total hypoxic range (air to PMMA) within the scale of 0–100 oxygen shortfall units (osu) [60].

One method of determining the response of the cornea to contact lens transmissibility is to measure corneal oxygen uptake rates immediately following 300s of lens wear [42], [43], [44], [45], [46], [47], [48], [49], [50], [51], [52]. Lenses of the same permeability but different center thicknesses can be used to determine the response of the cornea to that lens material across several transmissibilities [1]. The purpose of the study to be presented here was to determine the effects of sequencing of contact lens thicknesses on the measurement of the oxygen uptake rate of human corneas.

2. Methods 

return to Article Outline

2.1. Subjects 

Measurements were made on the right corneas of 14 human subjects, 8 males and 6 females, aged 24–58 years (mean 30 years). None of the subjects were contact lens wearers and all had good ocular health. Flattest keratometry values ranged from 41.25 to 44.37D (mean 43.25D), and corneas had 0.12–1.87D of corneal toricity (mean 1.08D). All subjects participated in at least three instructional and practice sessions and signed an informed consent form, reviewed and approved by the Institutional Review Board, the Biomedical Sciences Human Subjects Review Committee of The Ohio State University.

2.2. Contact lenses 

The contact lens materials that were used were polymethylmethacrylate and Siflufocon A, a fluorosilicone acrylate rigid gas permeable material with a Dk of 33×10−11cm2/s (mlO2)/(mlmmHg), a modulus of 1730MPa, a specific gravity of 1.25, and a silicone content of 7–8%. Both lens materials were manufactured to have overall diameters of 8.8mm, optic zone diameters of 7.4mm, axial edge lifts of 0.09mm, and back vertex powers of −3.00D. They were fitted “on K” (to match the radius of the flattest keratometer reading for each cornea). They were manufactured in seven center thickness: 0.08, 0.12, 0.16, 0.20, 0.24, 0.28, and 0.32mm. All lenses were verified to assure that all parameters were within ANSI standards and center thicknesses were within 0.01mm of the ordered values.

2.3. Procedure 

Oxygen uptake rates were measured on the central, non-anesthetized corneas of the right eyes of the 14 subjects. The equipment included a Clark-type polarographic electrode with a 25μm platinum cathode, fitted with a 12μm thick polyethylene membrane. Calibration was performed by alternately placing the electrode in saline baths saturated by air (155mmHg O2) and nitrogen (0mmHg O2), while maintained at 36°C.

Each subject participated in two measurement sessions. Oxygen uptake rates were measured for: (1) the normal open eye, (2) after 5min of static (without blinking) wear of each of the seven rigid gas permeable lenses of the seven center thicknesses, and (3) after 5min of static wear of the PMMA contact lens. Contact lenses were worn statically only, with no blinking permitted, so that corneal oxygenation was provided only through contact lens transmissibility and not through tear exchange with blinking. Subjects held their right upper lids, fixated a target with their eyes in the straight ahead position, and blotted any tears that accumulated in the outer and inner canthi during the 5min of lens wear. This prevented lens movement due to excessive tearing or eye movement. The first measurement of each session was that of the normal open eye. The second measurement of each session was after 5min of wear of the PMMA contact lenses. Five minutes with no contact lens wear and normal blinking ensued between lens applications.

The sequencing of the lens thicknesses for each subject was as follows. The lenses were randomly assigned a letter label (e.g., A, B, C, D, E, F, G). The lens code for each of the seven lens thicknesses can be seen in Table 1. Seven cycles lens sequences were determined (e.g., A, B, C, D, E, F, G; B C, D, E, F, G, A; C, D, E, F, G, A, B; D, E, F, G, A, B, C; E, F, G, A, B, C, D; F, B, A, B, C, D, E, F; G, A, B, C, D, E, F). Two subjects were assigned to each sequence. Table 2 shows the lens sequence assignments for each of the 14 subjects. Each subject participated in two identical sessions (with identical lens sequences), and data from the two sessions were averaged. Subjects were randomly assigned to the sequencing. The examiner was masked to the lens thicknesses and sequencing.

Table 1.

Lens code for each of the seven lens thicknesses

Lens codeLens thickness
A0.24
B0.20
C0.12
D0.28
E0.16
F0.32
G0.08
Table 2.

Lens sequence assignment for each of the 14 subjects

Subject numberSequence numberLens sequence
110.24, 0.20, 0.12, 0.28,0.16, 0.32, 0.08
210.24, 0.20, 0.12,0.28, 0.16, 0.32, 0.08
320.20, 0.12, 0.28, 0.16, 0.32, 0.08, 0.24
420.20, 0.12, 0.28, 0.16, 0.32, 0.08, 0.24
530.12, 0.28, 0.16, 0.32, 0.08, 0.24, 0.20
630.12, 0.28, 0.16, 0.32, 0.08, 0.24, 0.20
740.28, 0.16, 0.32, 0.08, 0.24, 0.20, 0.12
840.28, 0.16, 0.32, 0.08, 0.24, 0.20, 0.12
950.16, 0.32, 0.08, 0.24, 0.20, 0.12, 0.28
1050.16, 0.32, 0.08, 0.24, 0.20, 0.12, 0.28
1160.32, 0.08, 0.24, 0.20, 0.12, 0.28, 0.16
1260.32, 0.08, 0.24, 0.20, 0.12, 0.28, 0.16
1370.08, 0.24, 0.20, 0.12, 0.28, 0.16, 0.32
1470.08, 0.24, 0.20, 0.12, 0.28, 0.16, 0.32

A relative scale of 0–100 oxygen shortfall units can be used to rate human corneal responses to lens transmissibility. In this scale, 0osu represents the oxygen uptake rate of the normal open eye and 100osu represents the oxygen uptake rate associated with severe hypoxic stress (i.e., after an impermeable, un-pumped polymethylmethacrylate contact lens has been worn until maximum corneal oxygen uptake rates are reached).

where TL, oxygen uptake rate associated with the test lens; air, oxygen uptake rate associated with the normal, open eye; PMMA, oxygen uptake rate associated with PMMA reference lens.

Past studies in this laboratory have shown that corneal oxygen uptake rates for the normal open eye are about 5mmHg/s, while those following 5min of PMMA wear are about 30mmHg/s, with considerable inter-individual variation [42], [43], [44], [45], [46], [47], [48], [49], [50], [51], [52].

3. Statistical analyses 

return to Article Outline

Repeated measures analysis of variance was used to look at the effects on corneal oxygen shortfall of lens thickness, lens order, and the interaction between lens thickness and lens order. For each subject, the average of his or her two sessions (replications) is used to quantify the response at a given lens thickness.

4. Results 

return to Article Outline

Table 3 shows the oxygen response of the cornea, in osu, to seven sequences of seven lens thicknesses. Each value is a mean of two measurements on each of 14 subjects. Fig. 1 displays the mean corneal oxygen shortfall for the 14 subjects. The oxygen shortfall for each lens thickness sequence is displayed at each lens thickness.

Table 3.

Mean (S.D.) oxygen response of the cornea, in osu, to seven sequences of seven lens thicknesses

Thickness (mm)Lens sequenceOverall
1234567
0.0814.45 (2.95), n=210.62 (5.09), n=25.05 (3.50), n=210.40 (5.56), n=212.33 (4.82), n=215.41 (2.64), n=223.82 (4.31), n=213.15 (6.37), n=14
0.1262.09 (41.81), n=222.12 (1.96), n=265.85 (60.27), n=213.14 (4.61), n=236.80 (18.15), n=219.69 (1.11), n=221.78 (7.42), n=234.50 (29.44), n=14
0.1627.81 (2.77), n=226.62 (9.73), n=220.56 (6.41), n=234.68 (4.77), n=227.32 (13.06), n=221.46 (3.39), n=224.39 (13.98), n=226.12 (7.90), n=14
0.2036.96 (6.31), n=242.37 (18.64), n=221.88 (13.14), n=221.64 (0.98), n=226.40 (7.74), n=227.86 (5.07), n=241.84 (6.40), n=231.28 (11.28), n=14
0.2436.78 (10.25), n=217.06 (4.70), n=221.62 (8.03), n=225.40 (7.29), n=233.22 (16.62), n=230.87 (2.14), n=240.03 (18.42), n=229.28 (11.41), n=14
0.2842.59 (16.93), n=242.28 (6.58), n=235.32 (1.30), n=246.69 (30.45), n=228.46 (11.01), n=226.29 (1.19), n=236.36 (5.37), n=236.85 (12.70), n=14
0.3233.53 (6.21), n=225.53 (10.90), n=224.53 (6.42), n=227.87 (8.86), n=233.06 (15.88), n=231.80 (5.38), n=228.07 (29.07), n=229.20 (10.96), n=14
Overall36.32 (19.13), n=1426.66 (13.53), n=1427.83 (25.24), n=1425.69 (15.20), n=1428.23 (12.37), n=1424.77 (6.36), n=1430.90 (13.46), n=1428.63 (15.92), n=98

View full-size image.

Fig. 1. Mean oxygen shortfall of the human cornea associated with seven thicknesses (0.08–0.32mm in 0.04mm steps) of a single rigid gas permeable contact lens material, measured for seven thickness sequences.


The interaction of order x thickness was determine to be insignificant, using repeated measures ANOVA (F=0.99; p=0.51). The main effect of order was judged to be insignificant (F=0.76; p=0.63), indicating that the order of lens placement did not affect the osu value. The analysis indicated a significant effect of lens thickness on osu (F=3.94; p=0.003). Post hoc testing with an adjustment for multiple comparisons (Tukey's test, α=0.05) found that osu values obtained with the 0.12, 0.20, and 0.28mm lenses were significantly higher than those obtained with the 0.08mm thickness lens.

Two aberrant cell means stood out in an interaction plot used to test the interaction of order x thickness, allowing unstructured covariance matrix. In this plot, seven lines of cells means (mean osu for the two subjects at each order-thickness combination) were plotted against the thickness. The lines should be roughly parallel if there is no interaction. Both aberrant cell means involved unusually high oxygen shortfall associated with the 0.12mm thickness lens. This was caused by the data from two subjects, one with lens sequence 1 and the other with lens sequence 3. Both subjects had relatively low oxygen uptake rates associated with the wear of the PMMA contact lens in conjunction with relatively high oxygen uptake rates associated with the 0.12mm lens. The low oxygen uptake rates following PMMA lens wear was most likely due to a delay in polarographic electrode placement following lens removal. Tearing, lens movement, or lens decentration during the 5min wearing period could also result in reduced measured rates. On the other hand, rapid probe placement following the wear of the 0.12mm thick lens, or too much pressure between the polarographic electrode and the cornea, could have resulted in unrepresentatively high oxygen uptake rates associated with the 0.12mm lens.

5. Discussion 

return to Article Outline

In this paper, we looked at the effects of rigid gas permeable contact lens thickness sequencing on corneal oxygen uptake. Data were relativized as a proportion of the cornea's total hypoxic range and expressed in osu. It was determined that the sequencing of lenses of various thicknesses does not affect the measurement of corneal oxygen shortfall. That means that thick lenses are no more likely to affect subsequent measures of corneal oxygen uptake than are thin lenses. This result was expected because a 5-min period of recovery was provided between each period of contact lens wear. Prior studies have shown that, following the wear of PMMA contact lenses, recovery of normal oxygen uptake rates occurs within 60s of re-exposure to the normal air environment [4], [16], [61], [62]. The lenses used in this study were gas permeable contact lenses, which were associated with corneal oxygen uptake rates that were lower (less changed from baseline) than those associated with the wear of PMMA contact lenses. Past studies have shown that referencing data obtained under test lens conditions to that obtained for the normal open eye reduces differences due to instrument or corneal instabilities [63]. In future studies of the effects of contact lens transmissibility on corneal oxygen uptake, contact lens thicknesses can be used in random order.

References 

return to Article Outline

[1]. [1]Ostrem ED, Fink BA, Hill RM. Contact lens transmissibility: effects on delivery of oxygen to the cornea. Optom Vis Sci. 1996;73(3):159–163. MEDLINE | CrossRef

[2]. [2]Smith BJ, Fink BA, Hill RM. Corneal responses to lens transmissibility. J Am Optom Assoc. 1997;68(8):478–482. MEDLINE

[3]. [3]Larke JR, Parrish ST, Wigham CG. Apparent human corneal oxygen uptake rate. Am J Optom Physiol Opt. 1981;58(10):803–805. MEDLINE

[4]. [4]Hill RM, Fatt I. Oxygen deprivation of the cornea by contact lenses and lid closure. Am J Optom Physiol Opt. 1964;41(11):678–687.

[5]. [5]Hill RM, Fatt I. How dependent is the cornea on the atmosphere?. J Am Optom Assoc. 1964;35(10):873–875.

[6]. [6]Hill RM, Fatt I. Oxygen uptake from a reservoir of limited volume by the human cornea in vivo. Science. 1963;142:1295–1297. MEDLINE

[7]. [7]Hill RM, Fatt I. Oxygen measurements under a contact lens. Am J Optom Arch Am Acad Optom. 1964;41:382–387. MEDLINE

[8]. [8]Fatt I. Polarographic oxygen sensors. Cleveland, Ohio: CRC Press, Inc.; 1976;p. 1–7, 13–17, 25–33, 218–28.

[9]. [9]Fatt I. Measurement of oxygen flux into cornea by pressing a sensor onto a soft contact lens on the eye. Am J Optom Physiol Opt. 1978;55(5):294–301. MEDLINE

[10]. [10]Jaureghi MJ, Fatt I. Estimation of oxygen tension under a contact lens. Am J Optom Arch Am Acad Optom. 1971;48(3):210–218. MEDLINE

[11]. [11]Jaureghi MJ, Fatt I. Estimation of the in vivo oxygen consumption rate of the human corneal epithelium. Am J Optom Arch Am Acad Optom. 1972;49(5):507–511. MEDLINE

[12]. [12]Quinn T. Aerobic requirements of the human cornea, in College of Optometry. Columbus, Ohio: The Ohio State University; 1981;.

[13]. [13]Quinn TG, Schoessler JP. Solubility effects on corneal oxygen measurement. Am J Optom Physiol Opt. 1983;60(5):360–363. MEDLINE

[14]. [14]Benjamin WJ, Hill RM. Human cornea: oxygen uptake immediately following graded deprivation. Graefes Arch Clin Exp Ophthalmol. 1985;223(1):47–49. CrossRef

[15]. [15]Barber A, Fink BA, Hill RM. Human cornea: its rapid response to anoxia—the first 540 seconds. Invest Ophthalmol Vis Sci. 2000;77(12s):170;[ARVO Abstract].

[16]. [16]Fink BA, Smith BJ, Hill RM. Oxygen depletion characteristics of in vivo human cornea. Invest Ophthalmol Vis Sci. 1995;36(4):S310;[ARVO Abstract].

[17]. [17]Carney LG, Brennan NA. Time course of corneal oxygen uptake during contact lens wear. CLAO J. 1988;14(3):151–154. MEDLINE

[18]. [18]Holden BA, et al. Effects of long-term extended contact lens wear on the human cornea. Invest Ophthalmol Vis Sci. 1985;26(11):1489–1501. MEDLINE

[19]. [19]Paugh J. Physiological responses of the cornea to hydrogel lens parameters, in College of Optometry. Columbus, Ohio: The Ohio State University; 1981;p. 27.

[20]. [20]Hermann H, Moses SG, Friedenwald JS. Influence of pontacaine hydrochloride and chlorbutanol on respiration and glycolysis of cornea. Arch Ophthalmol. 1942;28.

[21]. [21]Bentley CR, Larke JR. Apparent respiration rate of the human corneal epithelium with tetracaine HCl and benoxinate HCl. Am J Optom Physiol Opt. 1983;60(12):960–963. MEDLINE

[22]. [22]Augsburger AR, Hill RM. Corneal anesthetics and epithelial oxygen flux. Arch Ophthalmol. 1972;88(3):305–307. MEDLINE

[23]. [23]Parrish ST, Larke JR. Apparent oxygen uptake rate of the human cornea in vivo following soft contact lens removal. Am J Optom Physiol Opt. 1981;58(9):696–698. MEDLINE

[24]. [24]Benjamin WJ. Corneal physiology under the closed eyelid of humans, in College of Optometry. Columbus, Ohio: The Ohio State University; 1982;.

[25]. [25]Benjamin WJ, Hill RM. Closed-lid factors influencing human corneal oxygen demand. Acta Ophthalmol (Copenh). 1986;64(6):644–648.

[26]. [26]Hill RM, Schoessler J, Fatt I. Oxygen flux across the environmental interface of teleost cornea. Nature. 1967;214(86):384–385. MEDLINE | CrossRef

[27]. [27]Benjamin WJ, Hill RM. Human corneal oxygen demand: the closed-eye interval. Graefes Arch Clin Exp Ophthalmol. 1986;224(3):291–294. CrossRef

[28]. [28]Fink BA, Carney LG, Hill RM. Influence of palpebral aperture height on tear pump efficiency. Optom Vis Sci. 1990;67(4):287–290. MEDLINE

[29]. [29]Heilman KE. The effect of the closed eye environment on the corneal oxygen shortfall associated with the wear of rigid gas permeable lenses, in College of Optometry. Columbus, Ohio: The Ohio State University; 1995;.

[30]. [30]Benjamin WJ, Hill RM. Human cornea: superior and central oxygen demands. Graefes Arch Clin Exp Ophthalmol. 1988;226(1):41–44. CrossRef

[31]. [31]Brunstetter TJ, Fink BA, Hill RM. What is the oxygen environment under an encapsulated segment bifocal RGP contact lens?. J Am Optom Assoc. 1999;70(10):641–646. MEDLINE

[32]. [32]Szczotka LB, Fink BA, Hill RM. Effect of prism ballasting of rigid contact lenses on human corneal oxygen uptake rates. Part I. Open eye conditions. ICLC. 1993;20:155–159.

[33]. [33]Szczotka LB, Fink BA, Hill RM. Effects of prism ballasting of rigid contact lenses on human corneal oxygen uptake rates. Part II. Closed-eye conditions. ICLC. 1994;21:112–126.

[34]. [34]Hill RM. Respiratory profiles of the corneal epithelium. I. Control profiles and effects of the non-aperture lens. Am J Optom Arch Am Acad Optom. 1966;43(4):233–237.

[35]. [35]Efron N, Fitzgerald JP. Distribution of oxygen across the surface of the human cornea during soft contact lens wear. Optom Vis Sci. 1996;73(10):659–665. MEDLINE | CrossRef

[36]. [36]Fink BA, Hill RM, Schoessler JP. An optical technique to predict corneal hypoxic response. Invest Ophthalmol Vis Sci. 1994;35(4):1776;[ARVO Abstract].

[37]. [37]Fink BA, Carney LG, Hill RM. Responses to oxygen deprivation: variations among human corneas. Graefes Arch Clin Exp Ophthalmol. 1991;229(3):287–290. CrossRef

[38]. [38]Hill RM, Schoessler J. Epithelial edema: respiratory characteristics of the lesion. Am J Optom Arch Am Acad Optom. 1968;45(4):241–245. MEDLINE

[39]. [39]Mauger TF. Aerobic responses of the cornea to alkali, in College of Optometry. Columbus, Ohio: The Ohio State University; 1981;.

[40]. [40]Roseman MJ. Aerobic responses of the cornea to isopropyl alcohol, measured in vivo, in College of Optometry. Columbus, Ohio: The Ohio State University; 1986;.

[41]. [41]Hill RM, Lowther GE. Oxygen flux across the tear-epithelial interface as an index of corneal wound repair. Am J Optom Arch Am Acad Optom. 1967;44(5):267–275. MEDLINE

[42]. [42]Fink BA. Quantifying the effects of contact lens prescription parameters on human corneal oxygen uptake, in College of Optometry. Columbus, Ohio: The Ohio State University; 1987;.

[43]. [43]Fink BA, Smith BJ, Hill RM, Jones LA. Effects of rigid contact lens truncation on oxygen access to the central cornea under static and dynamic conditions. ICLC. 1998;25:166–169.

[44]. [44]Fink BA, Carney LG, Hill RM. Rigid lens tear pump efficiency: effects of overall diameter/base curve combinations. Optom Vis Sci. 1991;68(4):309–313. MEDLINE | CrossRef

[45]. [45]Fink BA, Carney LG, Hill RM. Rigid contact lens design: effects of overall diameter changes on tear pump efficiency. Optom Vis Sci. 1991;68(3):198–203. MEDLINE | CrossRef

[46]. [46]Fink BA, Carney LG, Hill RM. Influence of power changes in single cut rigid contact lenses on tear pump efficiency. Optom Vis Sci. 1992;69(9):691–697. MEDLINE | CrossRef

[47]. [47]Fink BA, Carney LG, Hill RM. Influence of lens design on the tear pump efficiency associated with plus-power rigid contact lenses: a comparison of single cut, lenticular, and minus-carrier designs. Optom Vis Sci. 1993;70(1):2–8. MEDLINE | CrossRef

[48]. [48]Fink BA, Hill RM, Carney LG. Influence of rigid contact lens overall and optic zone diameters on tear pump efficiency. Optom Vis Sci. 1990;67(8):641–644. MEDLINE | CrossRef

[49]. [49]Fink BA, Hill RM, Carney LG. Effects of rigid contact lens edge lift changes on tear pump efficiency. Optom Vis Sci. 1991;68(6):409–413. MEDLINE | CrossRef

[50]. [50]Fink BA, Hill RM, Carney LG. Influence of rigid contact lens base curve radius on tear pump efficiency. Optom Vis Sci. 1992;69(1):60–65. MEDLINE | CrossRef

[51]. [51]Fink BA, Hill RM, Jones LA. Effects of rigid contact lens prism ballasting on oxygen access to the cornea under static and dynamic conditions. Optom Vis Sci. 1998;75(10):763–771. MEDLINE | CrossRef

[52]. [52]Fink BA, Hill RM, Carney LG. Rigid contact lens design equivalencies of overall diameter and axial edge lift: individual variations. Optom Vis Sci. 1993;70(9):733–738. MEDLINE | CrossRef

[53]. [53]Gardner HP, et al. The effects of high-Dk rigid contact lens center thickness, material permeability, and blinking on the oxygen uptake of the human cornea. Optom Vis Sci. 2005;82(6):459–466. MEDLINE | CrossRef

[54]. [54]Paugh JR. Equivalent oxygen percentage as a function of hydration in hydrogel lenses: an in vivo study. Optom Vis Sci. 1992;69(10):805–810. MEDLINE | CrossRef

[55]. [55]Berger RE. Effect of contact lens motion on the oxygen tension distribution under the lens. Am J Optom Physiol Opt. 1974;51(7):441–456. MEDLINE

[56]. [56]Young MD, Benjamin WJ. Calibrated oxygen permeability of 35 conventional hydrogel materials and correlation with water content. Eye Contact Lens. 2003;29(2):126–133. MEDLINE | CrossRef

[57]. [57]Efron N, Carney LG. Effect of blinking on the level of oxygen beneath hard and soft gas-permeable contact lenses. J Am Optom Assoc. 1983;54(3):229–234. MEDLINE

[58]. [58]Efron N, Carney LG. Models of oxygen performance for static, dynamic, and closed-lid wear of hydrogel contact lenses. Aust J Optom. 1981;64(6):223–233.

[59]. [59]Hill RM, Andrasko G. Oxygen and water. J Am Optom Assoc. 1981;52(3):225–226. MEDLINE

[60]. [60]Fink BA, Smith BJ, Hill RM. The human cornea's rapid response to hypoxia. Optom Vis Sci. 1999;76(12s):225.

[61]. [61]Benjamin WJ, Hill RM. Human cornea: individual responses to hypoxic environments. Graefes Arch Clin Exp Ophthalmol. 1988;226(1):45–48. CrossRef

[62]. [62]Hill RM. Oxygen uptake of the cornea following contact lens removal. J Am Optom Assoc. 1965;36(10):913–915. MEDLINE

[63]. [63]Fink BA, Carney LG, Hill RM. Intrasubject variability of human corneal oxygen uptake. ICLC. 1990;17:224–227.

The Ohio State University, College of Optometry, 338 West Tenth Avenue, Columbus, 43210-1240 OH, USA

Corresponding Author InformationCorresponding author. Tel.: +1 614 292 0084; fax: +1 614 292 7493.

PII: S1367-0484(06)00136-6

doi:10.1016/j.clae.2006.10.002